2-8 Preparing for Operation
Installing the reservoir bottles
Use standard reservoir bottles (15-mL) for instrument setup and calibration.
The Low-volume Adaptor Kit is included for infusing smaller volumes. The
capacity of the low-volume vials is 1.5 mL.
Use standard reservoir bottles (15-mL) for instrument setup and calibration.
Required material
Chemical-resistant, powder-free gloves
To install the reservoir bottles
1. Remove the reservoir bottle caps.
2. Screw the reservoir bottles onto the instrument, as shown below.
Warning: The reservoir bottles can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.