B-18 External Connections
Connecting to the nitrogen gas supply
Required materials:
Chemical-resistant, powder-free gloves
Sharp knife
•Wrench
6-mm PTFE tubing (included in the Waters Rough Pump Connect Kit)
To connect the nitrogen gas supply
1. Use the sharp knife to cut a 3.8 to 5.0-cm (1.5 to 2-inch) length of 6-mm
PTFE tubing.
2. Connect this piece of tubing to one end of the nitrogen supply in-line
filter.
3. Connect the remaining length of the 6-mm PTFE tubing to the other end
of the filter.
4. Connect the free end of the short piece of 6-mm PTFE tubing to the
nitrogen inlet port on the rear of the instrument.
Caution: To avoid gas leaks, use the sharp knife to cut the PTFE
tubing squarely (that is, perpendicular to its horizontal axis).