EasyManua.ls Logo

Waters TQ Detector - Replacing the APCI Probe Heater

Default Icon
235 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-72 Maintenance Procedures
Replacing the APCI probe heater
Replace the APCI probe heater if it fails to heat.
Required materials:
Chemical-resistant, powder-free gloves
Replacement probe heater
To replace the APCI probe heater
1. Remove the probe from the source (see “Removing the IonSABRE APCI
probe” on page 4-7).
2. Gripping the probe heater as shown, carefully pull it off the probe
assembly.
Warning: The probe and source components can be contaminated
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Caution: To avoid damaging the probe heater’s electrical
connections, do not twist the heater when removing it from the
probe assembly.
Probe heater connections

Table of Contents

Other manuals for Waters TQ Detector

Related product manuals