4-4 Optional APCI Mode of Operation
To install the IonSABRE APCI probe
1. Prepare the instrument for working on the source (see “Preparing the
instrument for work performed on the source” on page 5-7).
2. Open the instrument’s access door.
3. Ensure that the contacts on the IonSABRE APCI probe align with the
probe adjuster assembly contacts, and carefully slide the IonSABRE
APCI probe into the hole in the probe adjuster assembly.
Warning: The ACQUITY UPLC system connections, IonSABRE
APCI probe, and source can be contaminated with biohazardous
and/or toxic materials. Always wear chemical-resistant,
powder-free gloves while performing this procedure.
Warning: To avoid electric shock, ensure that the instrument is
suitably prepared before commencing this procedure.
Warning: The source can be hot. To avoid burn injuries, take
great care while working with the instrument’s access door open.
IonSABRE APCI
probe
Probe adjuster
assembly contacts
Probe adjuster assembly