Cleaning the source hexapole assembly 5-53
5. Use the 3-mm Allen wrench to fit and tighten the 3 PEEK ion block
support securing screws.
Tip: Look through the PEEK ion block support as you tighten the
securing screws. Ensure that neither hexapole spring buckles and
touches a hexapole rod.
6. Fit the ion block assembly to the PEEK ion block support (see “Fitting
the ion block assembly to the source assembly” on page 5-45).