5-100 Maintenance Procedures
6. Holding the oil demister element slightly elevated to prevent the loss of
the spring, remove the exhaust flange from the oil demister element.
7. Remove the spring from the oil demister element.
8. Dispose of the oil demister element in accordance with local
environmental regulations.
Warning: The oil demister element can be contaminated
with biohazardous and/or toxic materials. Ensure that it is
correctly disposed of according to local environmental
regulations.