Removing the ESI probe 3-11
7. Undo the 2 thumbscrews securing the ESI probe to the probe adjuster
assembly.
8. Carefully remove the ESI probe from the probe adjuster assembly.
9. If available, fit the protective sleeve to the ESI probe tip.
10. Close the instrument’s access door.
Warning: The ESI probe tip is sharp. To avoid puncture wounds,
handle the probe with care.