7351 Series Digital Multimeter Operation Manual
8.11 Error Message List
8-23
19 Err216 1/500 ATT measurement test error in analog operation
[ATT 1/500 Check]
Repair
20 Err500 Calibration data error
[CAL data]
Solution3
21 Err501 ZERO calibration is not performed.
[before ZERO]
Solution4
22 Err-102 Command syntax error
[Cmd Syntax]
Solution5
23 Err-113 The command cannot be used.
[Undefined header]
Solution5
24 Err-121 Unrecognized character as the numerical value
[Invalid character in number]
Solution5
25 Err-141 Invalid character string parameter is entered.
[Invalid character data]
Solution5
26 Err-151 Error as the character string data (Quotation mark error)
[Invalid string data]
Solution5
27 Err-200 Execution error (Command which cannot be executed now)
[Execution error]
Solution6
28 Err-213 The INITiate command is ignored because the measurement
is performing.
[Init ignored]
Solution6
29 Err-214 The READ? or MEAS? command is received when any trig-
ger source except for IMM is set.
[Trigger deadlock]
Solution6
30 Err-222 Entered value is outside the setting range or lacking for nec-
essary parameters
[Data out of range]
Solution6
31 Err-313 Calibration data is lost.
[Calibration memory lost]
Solution7
32 Err-314 Parameters saved by “*SAV” command are lost.
[Save/recall memory lost]
Solution8
33 Err-315 Saved parameters are lost.
[Configuration memory lost]
Solution6
34 Err-330 Self-test error
[Self-test failed]
Repair
35 Err-350 Error queue overflows.
[Queue overflow]
Solution9
No. Error display
Error description
[ ]: Response to the ERR? command
Solution