Chapter 2 Dispensing to SpectroCHIP Arrays30
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
To prime the pins:
1. Go to the Maintenance screen (tap the Back button on any screen and tap the
MAINTENANCE button on the Main Menu).
2. Tap the COMPLETE CYCLE button (Figure 2.12).
3. Select 5-10 complete cycles in the wash cycles screen that appears and tap OK
(Figure 2.13).
2.5 Setting up the dispensing run
Table 2.3 shows the recommended dispense settings. For most settings, it is not
recommended to change the recommended default value. More detail is provided in the
following sections.
Figure 2.12 Maintenance Screen
Figure 2.13 Wash Cycles Screen