EasyManuals Logo

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 User Manual

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000
162 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
Chapter 2 Dispensing to SpectroCHIP Arrays50
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
Checking the position
of each SpectroCHIP
Array
The position of each SpectroCHIP Array is checked to make sure that each is properly
aligned. The vision system locates the required fiducials (small dots on the SpectroCHIP
Array surface located between pads which are used for positioning and alignment) and
verifies that the SpectroCHIP Array is properly aligned. This video display may be viewed
by selecting the Vision tab on the Transfer screen. A warning message will appear if there
is a problem with SpectroCHIP Array positioning. See Appendix D for information on
troubleshooting SpectroCHIP Array positioning.
Reading MTP and
SpectroCHIP Array
barcodes
After the SpectroCHIP Array positions are checked, the SpectroCHIP Array and MTP
barcodes are read. Figure 2.46 shows an example of the video display when a
SpectroCHIP Array barcode is successfully read. This video display may be viewed by
selecting the Vision tab on the Transfer screen. A warning message will appear if there is
a problem with reading the barcodes. See Appendix D for information on troubleshooting
barcode reading.
Dispensing Dispensing begins after the SpectroCHIP Array positions are checked and barcodes are
read. Run activities can be monitored using the four tabs on the Transfer screen.
Transfer tab
The Transfer tab displays a graphic representation of the progress of dispensing from the
microtiter plate to the SpectroCHIP Array. Figure 2.47 shows the Transfer screen during
dispensing. Figure 2.48 shows the Transfer screen when dispensing is done.
Figure 2.46 Video Display Indicating
Successful Barcode Read

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 and is the answer not in the manual?

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 Specifications

General IconGeneral
BrandAgena BIOSCIENCE
ModelMassARRAY Nanodispenser RS1000
CategoryMeasuring Instruments
LanguageEnglish