3. To delete a user:
a. Tap the user name to be deleted (Figure 2.9).
b. Tap the DEL USER button in the User Setup screen.
c. Tap YES in the confirmation message that appears.
4. To edit user login information (for example, change the password or user level):
a. Delete the existing user (follow step 3).
b. Add a new user with the same login name, but new password and/or user level
(follow step 2).
2.4 Pre-run checks
Performing daily
maintenance
If you are the first person of the day to use the instrument, complete the daily
maintenance procedures. See Chapter 4, page 91 for daily maintenance instructions.
Checking the supply
and waste tanks
NOTENOTE
This section applies to the internal tanks. If you are using external tanks, see Checking the external
supply and waste tanks, page 141.
Before starting a dispensing run, confirm that the:
Supply tank is filled with Type 1, >18.2 MΩ water (a full tank is sufficient to process up
to ten 384 SpectroCHIP Arrays). See Chapter 4, page 99 for details on filling the
supply tank.
Waste tank is empty (an empty tank has the capacity to process ten 384 SpectroCHIP
Arrays). See Chapter 4, page 102 for details on draining the waste tank.
Tank status
If the STATUS button (at the top of most screens) is blinking, it means that the supply
tank is low and/or the waste tank is full.
Figure 2.9 Select and Confirm the User to be Deleted

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 and is the answer not in the manual?

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 Specifications

General IconGeneral
BrandAgena BIOSCIENCE
ModelMassARRAY Nanodispenser RS1000
CategoryMeasuring Instruments
LanguageEnglish