Chapter 2 Dispensing to SpectroCHIP Arrays 37
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
Selecting cleaning
parameters
Tap the cleaning tab on the Method screen. Verify that all cleaning steps are check-
marked in the Cleaning Setup box (Figure 2.22
).
Selecting aspirate/
dispense parameters
Tap the aspirate/dispense tab on the method screen.
To set the dispensing speed:
1. Under dispense settings, tap the dispense speed value (Figure 2.23). Enter a dispense
speed using the numeric keypad that appears, and tap enter.
• For iPLEX and MassCLEAVE, enter 100.
• For UltraSEEK, enter 120.
NOTENOTE
If your laboratory has determined optimal dispense speeds for your analytes that differ from the
recommended, use your values instead.
.
Figure 2.22 Cleaning tab
Figure 2.23 Set the Dispense Speed
Tap and enter a
dispense speed
value.