Chapter 3 Methods and Mappings70
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
3. Set the time for the cleaning steps:
a. To change a value, tap it and enter a new value in the number pad that appears.
b. Tap enter.
WARNING
!
WARNING! Do not set the wash time longer than 10 seconds. Longer times can damage the
ultrasonic wash.
Specify the aspirate and dispense settings
1. Tap the aspirate/dispense tab of the Method screen (Figure 3.17).
2. Choose the type of dispensing operation:
analyte only – Dispenses only analyte from microtiter plates to SpectroCHIP Arrays.
calibrant only – Dispenses only calibrant from the calibrant reservoir to the calibrant
pads on the SpectroCHIP Arrays.
analyte and calibrant – Dispenses both analyte and calibrant to the SpectroCHIP
Arrays.
Figure 3.16 Set Cleaning Times
Figure 3.17 Method Screen – Aspirate/Dispense Tab
These settings control
the duration of the
cleaning steps.
To change a
numeric value, tap
the value. In the
number pad that
appears, enter a
new value and tap
enter.