Appendix C External Tanks 143
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
C.6 Cleaning and flushing the wash system (for external tanks only)
NOTE
These instructions are for a setup using an external supply tank and an external waste tank. If you
are using a combination of internal and external tanks, for example, an internal supply tank with
an external waste tank (or vice versa), please contact Agena Bioscience Customer Support for
assistance.
This section explains how to clean the supply and waste tanks with ethanol and then
flush the wash system with a 65% ethanol solution.
CAUTION
!
It is critical that the cleaning and flushing procedures be performed monthly to prevent microbial
growth in the wash system.
WARNING
!
WARNING! You will be handling ethanol in this procedure. For safe handling procedures, refer
to the ethanol MSDS provided by your supplier. At a minimum: 1) wear protective clothing
(laboratory gloves, eye protection, and laboratory coat), 2) have adequate laboratory ventilation,
3) keep away from ignition sources, 4) and have proper fire extinguishing devices/solutions readily
accessible.
Drain the external
waste tank
1. Empty the external waste tank by following the instructions on page 142.
Drain the internal
supply tank
1. Go to the Maintenance screen (tap the Back button on any screen and tap the
MAINTENANCE button on the Main Menu).
2. Tap SUPPLY TANK Drain on the Maintenance screen.
Instructions on valve operation appear (Figure C.9).
3. Open the main door and lift the front access panel to access the tank control valves.
(See page 10 for instructions on opening the front access panel.)
Figure C.9 Draining the Supply Tank

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 and is the answer not in the manual?

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 Specifications

General IconGeneral
BrandAgena BIOSCIENCE
ModelMassARRAY Nanodispenser RS1000
CategoryMeasuring Instruments
LanguageEnglish