EasyManuals Logo

Agilent Technologies EMC Series Programmer's Guide

Agilent Technologies EMC Series
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #198 background imageLoading...
Page #198 background image
198 Chapter 5
Language Reference
ABORt Subsystem
ABORt Subsystem
Abort
:ABORt
Restarts any sweep or measurement in progress and resets the sweep or trigger
system. A measurement refers to any of the measurements found in the
MEASURE
menu.
If :INITiate:CONTinuous is off (single measure), then
:INITiate:IMMediate will start a new single measurement.
If :INITiate:CONTinuous is on (continuous measure), a new continuous
measurement begins immediately.
The INITiate and TRIGger subsystems contain additional related commands.
Front Panel
Access:
Restart for continuous measurement mode

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals