Chapter 5 281
Language Reference
[:SENSe]:DEMod Subsection
Demodulation Control
[:SENSe]:DEMod:STATe OFF|ON|0|1
[:SENSe]:DEMod:STATe?
Turns demodulation on or off.
Factory Preset
and *RST: Off
Front Panel
Access:
Det/Demod, Demod, Off
Demod Time
[:SENSe]:DEMod:TIME <time>
[:SENSe]:DEMod:TIME?
Sets the time used for frequency domain demodulation.
Factory Preset
and *RST: 500 ms
Range: 2 ms to 100 s
Default Unit: seconds
Front Panel
Access:
Det/Demod, Demod, Demod Time
Demod View
[:SENSe]:DEMod:VIEW[:STATe] OFF|ON|0|1
[:SENSe]:DEMod:VIEW[:STATe]?
This command causes the demodulated signal to be displayed. If FM Demod is on,
then the display scales the y-axis in units of kHz. The scale/div is set with the
command :DISPlay:WINDow:TRACe:Y
[:SCALe]:PDIVision:FREQuency <freq> if FM Demod is on. If FM
Demod is on, then several functions are not available; these include: Log/Lin
(display is always in linear), Y-Axis Units, Marker Search functions, Normalize,
Display Line, Peak Excursion, and Peak Threshold. There is no effect when AM
demodulation is used (only applicable for FM demodulation).
Factory Preset
and *RST: Off
Remarks: This command is not available when Demod is set to Off.
Front Panel
Access:
Det/Demod, Demod, FM, Demod View

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals