EasyManuals Logo

Agilent Technologies EMC Series Programmer's Guide

Agilent Technologies EMC Series
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #236 background imageLoading...
Page #236 background image
236 Chapter 5
Language Reference
DISPlay Subsystem
Factory Preset
and *RST: The factory default is On. This parameter is persistent, which
means that it retains the setting previously selected, even
through a power cycle.
Front Panel
Access:
System, Time/Date, Time/Date On Off
Display Annotation Title Data
:DISPlay:ANNotation:TITLe:DATA <string>
:DISPlay:ANNotation:TITLe:DATA?
Enters the text that will be displayed in the user title area of the display.
Front Panel
Access:
Display, Title
Display, Title, Change Title
Display, Title, Clear Title
Turn the Entire Display On/Off
:DISPlay:ENABle OFF|ON|0|1
Turns the display on or off. Having the display turned off may increase repetitive
measurement rate.
The following key presses will turn display enable back on:
1. If in local, press any key
2. If in remote, press the local (system) key
3. If in local lockout, no key (the computer must either cancel local lockout, or
re-enable the display)
Factory Preset
and *RST: On
Turn the Full Screen Display On/Off
:DISPlay:MENU:STATe OFF|ON|0|1
:DISPlay:MENU:STATe?
Turns the full screen display mode on or off. Press
System, System to turn off full
screen mode.
History: Added with firmware revision A.08.00.

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals