Chapter 5 295
Language Reference
[:SENSe]:POWer Subsection
[:SENSe]:POWer Subsection
Enable/Disable QPD X10 Gain
[:SENSe]:POWer:QPGain[:STATe]ON|OFF|1|0
[:SENSe]:POWer:QPGain[:STATe]?
Sets the quasi peak (QP) gain state On or Off in the quasi peak detector (QPD)
board.
Factory Preset
and *RST: Off
Key Access: Det/Demod, EMI Detector, AV/QP Gain X1 X10
Input Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the input attenuator. This value is set at its auto value if input attenuation is set
to auto.
Factory Preset
and *RST: 10 dB
Range: EMC E7401A: 0 to 60 dB
EMC E7402A: 0 to 75 dB
EMC E7403A: 0 to 75 dB
EMC E7404A: 0 to 75 dB
EMC E7405A: 0 to 65 dB
Default Unit: dB
Front Panel
Access:
AMPLITUDE/Y Scale, Attenuation Auto Man
Input Port Attenuator Auto
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Select the input port attenuator range to be set either automatically or manually.
On Input attenuation is automatically set as determined by the Reference
Level Setting.
Off Input attenuation is manually set

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals