Chapter 5 215
Language Reference
CALCulate:LLINe Subsection
Control Limit Line Testing
:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1
:CALCulate:LLINe[1]|2:STATe?
Turns limit line testing on/off. The limit and margin will only be tested if they are
displayed. Use :CALCulate:LLINe[1]|2:DISPlay to turn on the display
of limit lines, and :CALCulate:LLINe[1]|2:MARGin:STATe to turn on
the display of margins. If margin and limit display are both turned off, limit test is
automatically turned off. Use :CALCulate:LLINe[1]|2:FAIL? to return the
state of pass or fail after limit line state has been turned on.
Factory Preset
and *RST: Off
Front Panel
Access:
Display, Limits, Limit 1|2, Limit On Off
Select the Type of Limit Line
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2:TYPE?
Sets a limit line to be either an upper or lower type limit line. An upper line will be
used as the maximum allowable value when comparing with the data. A lower
limit line defines the minimum value.
Factory Preset
and *RST: Upper; not affected by preset
Remarks: If a margin has already been set for this limit line, and this
command is used to change the limit type, then the margin value
is reset to 0 dB.
Front Panel
Access:
Display, Limits, Limit 1|2, Type Upper Lower

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals