EasyManuals Logo

Agilent Technologies EMC Series Programmer's Guide

Agilent Technologies EMC Series
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #293 background imageLoading...
Page #293 background image
Chapter 5 293
Language Reference
[:SENSe]:FREQuency Subsection
Stop Frequency
[:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
Set the stop frequency.
Factory Preset
and *RST: 1.06 GHz
Range: EMC E7401A: 80 MHz
1
to 1.58 GHz
EMC E7402A: 80 MHz
1
to 3.10 GHz
EMC E7403A: 80 MHz
1
to 6.78 GHz
EMC E7404A: 80 MHz
1
to 13.3 GHz
EMC E7405A: 80 MHz
1
to 27.0 GHz
Default Unit: Hz
Front Panel
Access:
FREQUENCY/Channel, Stop Freq
Frequency Synthesis Mode
[:SENSe]:FREQuency:SYNThesis 1|2|3
[:SENSe]:FREQuency:SYNThesis?
This command switches between two phase noise optimization modes. Mode 2
optimizes the analyzer for close-in phase noise. Mode 3 optimizes the analyzer for
tuning speed. Mode 1 is not recommended for new designs.
This command is available for the following models only:
E7402A, E7403A, E7404A, E7405A
Factory Preset
and *RST: 3
History: Added with firmware revision A.08.00.
Front Panel
Access:
AUTO COUPLE, PhNoise Opt
1. 10 Hz minimum in log sweep mode.

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals