Chapter 5 307
Language Reference
SOURce Subsystem
Front Panel
Access:
Source, Amptd Step Auto Man
Set the Source Sweep Power Range
:SOURce:POWer:SWEep <rel_ampl>
:SOURce:POWer:SWEep?
Specifies the range of power levels through which the source output will sweep.
Use :SOURce:POWer:STARt to set the power level at the start of the power
sweep. See also :SOURce:POWer:SPAN.
Factory Preset
and *RST: 0 dB
Range: 0 dB to 20 dB
Default Unit: dB
Front Panel
Access:
Source, Power Sweep On Off
Output Power Tracking
:SOURce:POWer:TRCKing <integer>
:SOURce:POWer:TRCKing?
Adjusts the tracking of the source output with the spectrum analyzer sweep in the
present resolution bandwidth.
Factory Preset
and *RST: This command is persistent. The term persistent means that the
command retains the setting previously selected, even through a
power cycle.
Range: Integer, 0 to 4095
Remarks: This command is not needed with the 1.5 GHz tracking
generator.
Front Panel
Access:
Source, Man Track Adj
Output Power Tracking Peak
:SOURce:POWer:TRCKing:PEAK
Automatically adjusts the tracking of the source output with the spectrum analyzer
sweep so that the power is maximized for the present resolution bandwidth.
Remarks:. This command is not applicable for the 1.5 GHz tracking
generator.
Front Panel
Access:.
Source, Tracking Peak

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals