230 Chapter 5
Language Reference
CALibration Subsystem
Automatic Alignment
:CALibration:AUTO OFF|ON|0|1
:CALibration:AUTO?
Turns the automatic alignment on and off. This is run continuously, at the
completion of each sweep.
Factory Preset
and *RST: On at power-up
Front Panel
Access:
System, Alignments, Auto Align, All
System, Alignments, Auto Align, All but RF
System, Alignments, Auto Align, Off
Return to the Default Alignment Data
:CALibration:DATA:DEFault
Initializes the alignment data to the factory defaults.
Front Panel
Access:
System, Alignments, Load Defaults
Align FM Demodulation
:CALibration:FMDemod
:CALibration:FMDemod?
Performs an alignment of the FM Demodulation board. The query form of this
command performs the alignment and returns zero if the alignment is successful.
Front Panel
Access:
System, Alignments, Align Now, FM Demod
Query the Internal or External Frequency Reference
:CALibration:FREQuency:REFerence?
This is a query only. It reports the location of where the instrument frequency
reference is generated.
Range: INT or EXT

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals