EasyManuals Logo

Agilent Technologies EMC Series Programmer's Guide

Agilent Technologies EMC Series
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #280 background imageLoading...
Page #280 background image
280 Chapter 5
Language Reference
[:SENSe]:DEMod Subsection
[:SENSe]:DEMod Subsection
Type of Demodulation
[:SENSe]:DEMod AM|FM
[:SENSe]:DEMod?
Sets the type of demodulation.
Factory Preset
and *RST: AM
Front Panel
Access:
Det/Demod, Demod, AM
Det/Demod, Demod, FM
FM Deviation
[:SENSe]:DEMod:FMDeviation <freq>
[:SENSe]:DEMod:FMDeviation?
Sets the total FM frequency deviation for full screen demodulation.
Factory Preset
and *RST: 100 kHz
Range: 5 kHz to 1.2 MHz
Default Unit: Hz
Front Panel
Access:
AMPLITUDE, Scale/Div
Squelch
[:SENSe]:DEMod:SQUelch <integer>
Sets the squelch level on FM demod.
Factory Preset
and *RST: integer, 0 to 100
Key Access:
Det/Demod, Demod/Audio, Squelch