214 Chapter 5
Language Reference
CALCulate:LLINe Subsection
Test the Data Against the Limit Line
:CALCulate:LLINe[1]|2:FAIL?
Queries the status of the limit line testing. Returns a 0 if the data passes, and
returns a 1 if there is a failure. This query value is valid only if margin or limit test
is On. Use the command :CALCulate:LLINe[1]|2:STATe OFF|ON|0|1 to
activate limit line testing.
Set the Margin Size
:CALCulate:LLINe[1]|2:MARGin <rel_ampl>
:CALCulate:LLINe[1]|2:MARGin?
Allows you to define the amount of measurement margin that is added to the
designated limit line.
Factory Preset
and *RST: not affected
Default Units: dB
Remarks: The margin must be negative for upper limit lines, and positive
for lower limits.
Front Panel
Access:
Display, Limits, Limit 1|2, Margin On Off
Display the Limit Margin
:CALCulate:LLINe[1]|2:MARGin:STATe OFF|ON|0|1
:CALCulate:LLINe[1]|2:MARGin:STATe?
Allows you to display a measurement margin that is added to the designated limit
line to do secondary testing of the data.
Factory Preset
and *RST: Off
Front Panel
Access:
Display, Limits, Limit 1|2, Margin On Off

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals