Chapter 5 231
Language Reference
CALibration Subsystem
Coarse Adjust the Frequency Reference
:CALibration:FREQuency:REFerence:COARse <setting>
:CALibration:FREQuency:REFerence:COARse?
Allows coarse adjustment of the internal 10 MHz reference oscillator timebase of
the analyzer.
NOTE :CALibration:ALL is required after COARse is set.
Range: Integer, 0 to 255
Front Panel
Access:
System, Alignments, Time Base, Coarse
Fine Adjust the Frequency Reference
:CALibration:FREQuency:REFerence:FINE <setting>
:CALibration:FREQuency:REFerence:FINE?
Allows fine adjustment of the analyzer internal 10 MHz reference oscillator
timebase.
NOTE :CALibration:ALL is required after FINE is set.
Range: Integer, 0 to 255
Front Panel
Access:
System, Alignments, Time Base, Fine
Select the Frequency Corrections
:CALibration:FREQuency[:STATe] OFF|ON|0|1
:CALibration:FREQuency[:STATe]?
Turns on or off the frequency corrections.
Factory Preset
and *RST: On
Front Panel
Access:
System, Alignments, Freq Correct On Off
Align the RF Circuitry
:CALibration:RF
:CALibration:RF?
Performs an alignment of the RF assembly.
The query performs the alignment and returns a zero if the alignment is successful.

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals