EasyManuals Logo

Agilent Technologies EMC Series Programmer's Guide

Agilent Technologies EMC Series
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #299 background imageLoading...
Page #299 background image
Chapter 5 299
Language Reference
[:SENSe]:SWEep Subsection
Factory Preset
and *RST: Linear
Remarks: Refer to the Users Guide for detailed information on the
interactions of this command with other functions.
History: Added with firmware revision A.08.00.
Front Panel
Access:
FREQUENCY, Scale Type
Sweep Time
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Specifies the time in which the instrument sweeps the display.
Factory Preset
and *RST: 120.8 ms
Range: The range depends upon the installed options, number of sweep
points, and firmware revision of your instrument. See Sweep
Time Range in the Specifications Guide for details.
Default Unit: seconds
Remarks: A span value of 0 Hz causes the analyzer to enter zero span
mode. In zero span the X-axis represents time rather than
frequency. In this mode, the sweep time may be set to faster
values when Option AYX is installed.
Front Panel
Access: Sweep, Sweep Time Auto Man
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time for the current settings.
Factory Preset
and *RST: On
History: This command function changed with firmware revision
A.08.00. With :AUTO ON in zero span, an error will be
generated.
Front Panel
Access:
Sweep, Sweep Time Auto Man

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals