Chapter 5 233
Language Reference
COUPle Subsystem
COUPle Subsystem
Some measurement settings are automatically coupled together to optimize speed
and accuracy. These commands control that coupling.
COUPle the Function to Other Settings
:COUPle ALL|NONE
:COUPle?
The instrument can automatically couple instrument settings together for accurate
measurements and optimum dynamic range. This command is used to override the
coupling for special measurement needs.
COUPle NONE puts these functions into the manually set (not coupled) mode.
COUPle ALL puts the functions into the auto coupled mode, and also puts the
sweep coupling mode into SA (couple all).
The following list of analyzer functions can be automatically coupled:
Resolution bandwidth
Center Frequency
Average type (Firmware revision A.08.00 or greater)
Marker functions
Detector (Firmware revision A.08.00 or greater)
Marker functions
Average On Off
Average type
Attenuation
Reference level
External amplifier gain
Preamp
Center frequency step
Span (in swept spans)
Resolution bandwidth (in zero spans)
Video bandwidth
Resolution bandwidth
Tracking Generator
Sweep coupling mode (SR/SA)
VBW/RBW ratio (Firmware revision A.08.00 or greater)
Sweep time
Span
Video bandwidth
Resolution bandwidth

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals