Chapter 5 297
Language Reference
[:SENSe]:POWer Subsection
Remarks: This command is available only on Agilent EMC models
E7403A, E7404A, and E7405A. Use this command for signals
close to the noise level, multiple signals close together, or for
other conditions when the preselector is not tuned to the
frequency of interest.
Front Panel
Access:
AMPLITUDE/Y Scale, Presel Adjust
Preselector Center
[:SENSe]:POWer[:RF]:PCENter
This command centers the preselector filter at the signal of interest. This command
has no effect if it is activated in non-preselected bands. This command is usable
from 3 GHz to the maximum frequency of the analyzer.
NOTE This command is available only on Agilent EMC models E7403A, E7404A, and
E7405A. This command has no effect with markers set to less than 3 GHz.
Remarks: A peak search will be done if no marker is on.
Front Panel
Access:
AMPLITUDE/Y Scale, Presel Center