328 Chapter 5
Language Reference
TRACe Subsystem
TRACe Subsystem
The TRACe subsystem controls access to the internal trace memory of the
analyzer.
NOTE Refer also to :CALCulate and :MMEMory subsystems for more trace and limit
line commands.
Copy Trace
:TRACe:COPY <source_trace>,<dest_trace>
Transfers the source trace to the destination trace and leaves the destination trace
in VIEW mode.
Source traces are: TRACE1|2|3
Destination traces are: TRACE1|2|3
Example: :TRAC:COPY TRACE2,TRACE1
Front Panel
Access:
View/Trace, Operations, 1 > 3
View/Trace, Operations, 2 > 3
Transfer Trace Data
:TRACe[:DATA] <trace_name>|RAWTRACE,<definite_length_
block>|<comma_separated_ASCII_data>
:TRACe[:DATA]? <trace_name> |RAWTRACE|LLINE1|LLINE2
This command transfers trace data from the controller to the instrument. The data
format is set by the command :FORMat [:TRACe][:DATA]. The data is
comma-separated ASCII values in ASCII formatting, and a definite length block in
REAL, INTeger, and UINTeger formatting.
The query returns the current values of the designated trace. The data is terminated
with <NL><END> (for GPIB that is newline, or linefeed, followed by EOI set
true; for RS-232 this is newline only.)
LLINE1 and LLINE2 can only be queried; they cannot be set.
<trace_name> is TRACE1|2|3

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals