EasyManuals Logo

Agilent Technologies EMC Series Programmer's Guide

Agilent Technologies EMC Series
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #257 background imageLoading...
Page #257 background image
Chapter 5 257
Language Reference
MEASure Group of Commands
Measurement settings persist if you initiate a different measurement and then
return to a previous one. Use READ:<measurement>? if you want to use those
persistent settings. If you want to go back to the default settings, use
MEASure:<measurement>?.
Get Measurement Results
:FETCH?
Returns the results of the last measurement.
Factory Preset
and *RST: Not affected by *RST or Preset
Remarks: Only returns data if a measurement has been made. Returns the
results of the measurement as a string.
Read Command
READ?
Performs configured measurement and returns the results in a string format.
Factory Preset
and *RST: N/A
Remarks: Output format for measurement results of a single signal from
the Measure at Marker or Measure at Frequency keys is:
<PEAK>, <QUASI-PEAK>, <AVG>, <FREQ
UNCERTAINTY>, <TOTAL AMPLITUDE CORRECTION>,
<COMMENT>
Peak and signal list measurement output is the number of
signals measured.
Configure for Measuring Frequency
:CONFigure:EMI:AFRequency <freq><unit>
:MEASure:EMI:AFRequency? <freq><unit>
Configures for/measures a given frequency.
Factory Preset
and *RST: Not affected by *RST or Preset.

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals