296 Chapter 5
Language Reference
[:SENSe]:POWer Subsection
Factory Preset
and *RST: On
Front Panel
Access:
AMPLITUDE Y Scale, Attenuation
Input Port Power Gain
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Turns the internal preamp on or off.
Factory Preset
and *RST: Off
Front Panel
Access: AMPLITUDE/Y Scale, Int Preamp On Off
Input Port Maximum Mixer Power
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <ampl>
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
Specifies the maximum power at the input mixer.
Factory Preset
and *RST: 10 dBm
Range: 100 dBm to 10 dBm
Default Unit: dBm
Front Panel
Access:
AMPLITUDE/Y Scale, Max Mixer Lvl
Optimize Preselector Frequency
[:SENSe]:POWer[:RF]:PADJust <freq>
[:SENSe]:POWer[:RF]:PADJust?
This command allows user-defined adjustment of the preselector frequency to
optimize its response on the signal of interest.
Factory Preset
and *RST: 0 Hz
Range: 250 MHz to 250 MHz
Default Unit: None. Use the MHz terminator in order for this command to
work.

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals