240 Chapter 5
Language Reference
DISPlay Subsystem
Trace Y-Axis Amplitude Scaling
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision <rel_ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision?
Sets the per-division display scaling for the y-axis when y-axis units are set to
amplitude units.
Factory Preset
and *RST: 10 dB
Range: 0.1 to 20.0 dB
Default Unit: dB
Front Panel
Access:
AMPLITUDE/Y Scale, Scale/Div
Trace Y-Axis Frequency Scaling
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision:FREQuency <freq>
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision:FREQuency?
This command sets the per-division display scaling for the y-axis, when the y-axis
units are set to frequency units, such as when looking at FM deviation with the
command [:SENSe]:DEMod:VIEW[:STATe] OFF|ON|0|1.
Factory Preset
and *RST: 20 kHz
Range: 1 kHz to 240 kHz
Default Unit: Hz
Front Panel
Access:
AMPLITUDE/Y Scale, Scale/Div
Trace Y-Axis Reference Level
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
Sets the amplitude value of the reference level for the y-axis.
Factory Preset
and *RST: 107 dBµV
Range: EMC E7401A: 149.9 to 50 dBm
EMC E7402A, E7403A, E7404A, E7405A:
149.9 to 55 dBm

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals