272 Chapter 5
Language Reference
[:SENSe]:BANDwidth Subsection
[:SENSe]:BANDwidth Subsection
Resolution Bandwidth
[:SENSe]:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
Specifies the resolution bandwidth.
Example: BAND 1 kHz
Range: 10 Hz to 5 MHz (with Option 1DR)
1 Hz to 5 MHz (E7401A, E7402A, E7403A, E7404A, E7405A
with Option 1D5)
Default Unit: Hz
Front Panel
Access:
BW/Avg, Resolution BW Auto Man
Resolution Bandwidth Automatic
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO?
Couples the resolution bandwidth.
Factory Preset
and *RST: On
Example: BWID:AUTO On
History: This command function changed with firmware revision
A.08.00. With :AUTO ON in zero span, an error will be
generated.
Remarks: Auto-couple resolution bandwidth is not available in zero span.
Resolution Bandwidth Mode
[:SENSe]:BANDwidth|BWIDth[:RESolution]:MODE EMI|SAN|OFF
[:SENSe]:BANDwidth|BWIDth[:RESolution]:MODE?
Couples the resolution bandwidth to the frequency span in SAN mode, to Center
Frequency in EMI mode, or uncoupled when OFF.
Factory Preset
and *RST: EMI

Table of Contents

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals