107
Repairing the Autosampler
Simple Repairs
Simple Repairs
The procedures described in this section can be done with the autosampler in
place in the stack. You will do some of these procedures on a more frequent
basis
.
Table 16 Simple Repair Procedures
Procedure Typical Frequency Time Required Notes
Exchanging the needle
assembly
When needle shows indication of
damage or blockage
15 minutes See Needle Assembly on page
108
Exchanging the seat
assembly
When the seat shows indication of
damage or blockage
10 minutes See Needle-Seat Assembly on
page 111
Exchanging the rotor
seal
After approximately 30000 to 40000
injections, or when the valve
performance shows indication of
leakage or wear
30 minutes See Rotor Seal on page 117
Exchanging the
metering seal
When autosampler reproducibility
indicates seal wear
30 minutes SeeMetering Seal and Plunger on
page 120
Exchanging the gripper
arm
When the gripper arm is defective 10 minutes SeeGripper Arm on page 123

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1313A and is the answer not in the manual?

Agilent Technologies G1313A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1313A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals