Troubleshooting and Test Functions
Restart Without Cover
Restart Without Cover
The autosampler was restarted with the top cover and foam open.
The sensor on the ASM board detects when the top foam is in place. If the
autosampler is restarted with the foam removed, the autosampler switches
off within 30 s, and the error message is generated.
Probable Causes autosampler started with the top cover and foam removed.
Suggested Actions Replace the top cover and foam.