Troubleshooting and Test Functions
Troubleshooting Guide for the Sample Transport Assembly
Troubleshooting Guide for the Sample
Transport Assembly
This troubleshooting guide is meant to help you diagnose and repair
autosampler problems.
In general, autosampler problems can be divided into three categories.
1 Intermittent lock-ups with or without vial in the gripper fingers
with error messages
motor overtemp (0 or 1 or 2 or 3)
movement failed (0 or 1 or 2 or 3)
missing vial
Many times the sampler is being used very heavily.
2 Jittery (shaky) movement in X and/or theta axes and/or when the
needle goes through the gripper arm into the vial with error
messages
motor overtemp (0 or 2)
movement failed (0 or 2)
3 Poor alignment, seen during vial pickup and vial replacement
and/or when the needle hits the gripper arm with error messages
motor overtemp (0 or 2 or 3)
movement failed (0 or 2 or 3)
missing vial
NOTE Motor 0=X; 1=Z; 2=Theta; 3=Gripper.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1313A and is the answer not in the manual?

Agilent Technologies G1313A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1313A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals