Troubleshooting and Test Functions
Safety Flap Missing
Safety Flap Missing
The safety flap was not detected.
Before the needle moves down into the needle seat to inject sample, the
safety flap locks into position. Next, and the gripper checks the safety flap by
trying to move the safety flap away from the needle. If the gripper is able to
move beyond the safety flap position (safety flap not in position), the error
message is generated.
Probable Causes Safety flap missing or broken.
Suggested Actions Exchange the safety flap.