Theory of Operation
Firmware Description
Firmware Description
The firmware of the instrument consists of two independent sections:
a non-instrument specific section, called ‘resident system’,
an instrument specific section, called ‘main system’.
Resident System
This resident section of the firmware is identical for all Agilent 1100 series
modules. Its properties are:
the complete communication capabilities (GPIB, CAN, LAN and
RS-232C),
memory management,
ability to update the firmware of the ‘main system’.
Main System
Its properties are:
the complete communication capabilities (GPIB, CAN, LAN and
RS-232C),
memory management,
ability to update the firmware of the ‘resident system’.
In addition the main system comprises the instrument functions that are
divided into common functions like
run synchronization via APG remote
error handling,
diagnostic functions, and so on,
or module specific functions like
internal events such as metering device, gripper and needle movements,