154
Repairing the Autosampler
Power Supply
Power Supply
CAUTION The ASM board is sensitive to electrostatic discharge. Always use the ESD
strap (see Using the ESD Strap on page 105) when handling electronic
boards.
Frequency When defective
Tools required Pozidrive No. 1 screwdriver
Parts required Power supply 0950-2528.
Before beginning this procedure:
Remove the top cover (page 128).
Remove the transport assembly (page 132).
Remove the sampling unit (page 134).
Remove the ASM board (page 148).
Remove the fan (page 146).
1 Remove the bottom foam.
2Remove the power switch light pipe from the
coupler.
3Remove the two screws from the rear panel.
CellFrame
CellFrame
Light
pipe
Coupler
CellFrame