Specifications
Performance Specifications
Performance Specifications
Table 62 Performance Specifications Agilent 1100 Series Autosampler (G1313A) and
Thermostatted Autosampler (G1329A). Valid when standard 100 µl metering head
installed.
7\SH 6SHFLILFDWLRQ
GLP features Early maintenance feedback (EMF), electronic records of
maintenance and errors
Communications Controller-area network (CAN). GPIB (IEEE-448),
RS232C, APG-remote standard, optional four external
contact closures and BCD vial number output
Safety features Leak detection and safe leak handling, low voltages in
maintenance areas, error detection and display
Injection range 0.1 100 µl in 0.1 µl increments
Up to 1500 µl with multiple draw (hardware modification
required)
Replicate injections 1 99 from one vial
Precision Typically < 0.5 % RSD of peak areas from 5 100 µl,
Typically <1%RSD of peak areas from 1–5µl
Minimum sample
volume
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from
10 µl sample in 300 µl microvial
Carryover Typically < 0.1 %, < 0.05 % with external needle cleaning
Sample viscosity range 0.2 50 cp
Replicate injections per
vial
1–99
Sample capacity 100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray
15 × 6-ml vials in ½ tray (Agilent vials only)
Injection cycle time Typically 50 s depending on draw speed and injection
volume

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1313A and is the answer not in the manual?

Agilent Technologies G1313A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1313A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals