Troubleshooting and Test Functions
Initialization with Vial
Initialization with Vial
The autosampler attempted to initialize with a vial still in the gripper.
During initialization, the autosampler checks correct operation of the gripper
by closing and opening the gripper fingers while monitoring the motor
encoder. If a vial is still in the gripper when initialization is started, the
gripper fingers cannot close causing the error message to be generated.
Probable Causes Vial still in gripper.
Suggested Actions Remove the vial using the Release Vial function in the user interface.
Reinitialize the autosampler.