EasyManua.ls Logo

Agilent Technologies G1313A - Page 45

Agilent Technologies G1313A
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
Optimizing Performance
Fast Injection Cycle and Low Delay Volume
low flow rates this can lead to retention time changes which are to big for the
actual calibration table. To overcome this it is recommended to add an
additional “blank” injection as last injection to the sequence.
General Recommendations for Fast Injection Cycle Times
As described in this section, the first step to provide short cycle times are
optimizing the chromatographic conditions. If this is done the autosampler
parameter should be set to:
• Overlapped injection mode
• Increase of draw and eject speed for large injection volumes
• Add at last run a blank, if overlapped injection is used
To reduce the injection time, the detector balance has to be set to OFF.

Table of Contents