7-34 i3070 Series 5i Help
10 Tighten the locknuts and the screw on the coupling.
Reposition Set Collars
11 Loosen the screws on the set collars and move them back into position,
30 mm to 50 mm from the end of the shaft. Tighten the screws on the
set collars.
12 Close all doors.
Locknuts
Screw