5-2 i3070 Series 5i Help
Maintenance
Calibration
All Medalist i3070 ICT systems have been tested and calibrated by Agilent 
Technologies to ensure that all warranted specifications are met at the 
time of shipment.
For ongoing calibration and performance verification:
• ASRU Calibration – Agilent recommends that the ASRU Card in every 
module of a Medalist i3070 ICT system be calibrated once every six 
months. 
This procedure calibrates several voltage references and a resistance 
reference on the ASRU Card. These references are then used to 
automatically derive all other AutoAdjust correction constants. The 
system Diagnostics and AutoAdjust routines constitute an overall system 
operational verification. The ASRU calibration is the only calibration 
required for the system to continue performing at system specifications.
• System Calibration – In addition to the ASRU calibration, a System 
Calibration Service is available for customers who desire more 
comprehensive measurements and reporting. System Calibration should 
be performed every six months. 
This service uses a BT- BASIC program and a traceable voltmeter and 
frequency counter to measure a DC voltage source at 20 voltages, an AC 
voltage source at nine voltages, a resistor, and two oscillator 
frequencies.
These measurements are the foundation for all analog measurements 
and digital timing. These same parameters are also measured using the 
internal detectors in the testhead, and the measurements are compared 
to the standards set by the voltmeter and frequency counter to calibrate 
the detectors. A report is generated containing the measurement and 
uncertainty data.
A Pin Verification Fixture (PVF, part number E9988- 67901) is required for 
the calibration procedure. For detailed information about the requirements 
and calibration procedures, refer to Calibration in the Medalist i3070 
online help.