i3070 Series 5i Help 7-47
3 To remove the spring from the fixed rail assembly:
a Remove the screw securing the spring guide from below.
b Remove four screws securing the support plate.
The support plate, spring, and spring guide can now be removed.
Installation
1 Assemble the new spring, spring guide, and support plate.
Secure the support plate to the rail assembly with four screws.
2 Secure the spring guide with one screw from below the assembly.
3 Install the fixed rail assembly in the system, securing it with two
screws on either side.
4 Close all covers/doors.
Spring Guide
Screw
Support Plate