Troubleshooting
i3070 Series 5i Help 6-9
3 Press Origin to move the Press to its start position.
4 If the lower limit sensor is too high:
Loosen the two screws securing the sensor bracket (Figure 6- 1) and
move the sensor down to allow the Press to move down.
5 Press and hold Down to lower the Press until all the probes are
contacting the board.
6 If the lower limit sensor is too low:
Loosen the two screws securing the sensor bracket (Figure 6- 1).
7 Adjust the position of the lower limit sensor:
a Slide the sensor slowly upwards until it is activated by the sensing
plate (the red sensor light will turn on).
b Slide the sensor down slightly, just enough to turn the sensor light
off. Tighten the two screws on the sensor bracket.
Press Slow to slow the movement of the Press if
necessary.
NOTE
Lower Limit Sensor
Sensing Plate
Sensor Light