PN 903-8797, Rev A
The Stat\History menu items (listed in their viewing order) and their descrip-
tions are:
Sub-Menu Item Description
En\Dis Alrm? This prompt asks if you wish to enable\disable the out-of-specication (O-Spec) alarm.
To enable the O-Spec Alarm, press +. To disable the O-Spec Alarm, press ’.
Clear Alarm? This prompt asks if you wish to clear (reset) all current alarms (O-Spec or SCode).
To reset all current alarms, press +. The message “Cleared will be displayed. If you press ’,
Not Cleared will be displayed.
SCode z Basic Status Code and messages, where ‘z’ = the basic SCode (see below) along with its
corresponding message (if the alarm condition exists). If the alarm condition does not exist,
OK is displayed (e.g., SCode.# OK, where ‘#’ is the number corresponding with the SCode
message).
The basic SCodes and corresponding messages are listed below. The ‘w or ‘f preceding a
message indicates a ‘w’arning or ‘f’ault alarm condition.
Primary
Code
Error
Code # Message [Description]
1 1 w TimeOut [Stage-Time-Out]
2 2 w No H2O DP [Water-Dewpoint-Temperature-Not-Found] –
Available only if H2O DP:INT option is enabled from the
MonitorCFG menu.
4 3 w No HC DP [Hydrocarbon-Dewpoint-Temperature-Not-Found]
8 4 w Check XSCode [Advice – Review-Extended-Status-Code]
16 5 w Dirty Cell [Possible-Cell-Contamination]
32 6 f H2O Optics [Problem-In-Water-Side-Optical-Path]
64 7 f HC Optics [Problem-In-Hydrocarbon-Side-Optical-Path]
128 8 f Cool Cct [Problem-In-Cooling-Circuit]
XSCode z Extended Status Codes and messages, where ‘z’ = the XSCode (see below) along with its
corresponding message (if the alarm condition exists). If the alarm condition does not exist,
OK is displayed (e.g., XSCode.# OK, where ‘#’ is the number corresponding with the XSCode
message).
The XSCodes and corresponding messages are listed below. The ‘w or ‘f preceding a message
indicates a ‘w’arning or ‘f’ault alarm condition.
Primary
Code
Error
Code # Message [Description]
1 1 w EEP Warn [End-Of-EEPROM-Life]
2 2 w Hold TimeOut [Purging/Hold-Stage-Time-Out]
4 3 w Cool TimeOut [Cooling-Stage-Time-Out]
8 4 w Warm TimeOut [Warming-Stage-Time-Out]
16 5 w Cross Talk [Excessive-Mirror-Cross-Talk]
32 6 w High DP [Hydrocarbon-Dewpoint-Temperature-Too-High]
64 7 f High Temp [Mirror-RTD-Temperature-Is-Too-High]
128 8 f Low Temp [Mirror RTD-Temperature-Is-Too-Low]
256 9 f H2O Dim [Light-Intensity-On-Water-Side-Is-Too-Low]
512 10 f HC Dim [Light-Intensity-On-Hydrocarbon-Side-Is-Too-Low]
3-12 | 241CE II Hydrocarbon Dewpoint Analyzer

Table of Contents

Other manuals for Ametek 241CE II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 241CE II and is the answer not in the manual?

Ametek 241CE II Specifications

General IconGeneral
BrandAmetek
Model241CE II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals