3. If everything appears to be set correctly in the analyzer’s memory, the next
step is to troubleshoot the sensor and heating circuitry. This will require
removing the analyzer and taking it to your shop bench.
4. Remove the four (4) screws that secure the lid and remove the four (4) cor-
ner screws securing the Electronics Enclosure to the Cell Oven Enclosure.
Take the Electronics Enclosure and have it lay to the side of the main black
box. It will only extend as far as the wire harness will allow; take care not to
pull or damage the wires.
5. Connect your computer to the analyzer and power up the computer, and
run the Congurator Software.
6. Apply the required 24 VDC power to the unit.