x | Model 900 ADA / Model 930 Sulfur Pit Analyzers
Because ultraviolet radiation can harm your eyes, avoid direct viewing of the light emanating
through the end window of the source lamp. If the source lamp must be viewed while ener-
gized, wear safety glasses that block ultraviolet radiation.
The Analyzer Oven enclosure and components within the Analyzer Oven are hot; take precau-
tions to avoid burning yourself.
Purged Analyzer (Hazardous Location) Applications
[Special Conditions for Safe Use]
The analyzer may only be energized by using the Purge Bypass Switch with permission of
the works manager or his proxy. The permission may only be given when it is made sure that
during the time the system is energized by using this switch an explosive atmosphere is not
present or when the necessary protective measures against explosion hazard have been taken
(“hot permit”).
The analyzer enclosure may not be opened when an explosive atmosphere is present.
Purged Analyzer (Hazardous Location) Applications
Do not apply power to the analyzer if there is damage (scratches, indentations, or wear) to any
flamepath (on the Oven Heater or Disconnect Enclosure). Applying power to an analyzer with
a damaged flamepath is dangerous and could result in serious injury or death, or serious dam-
age to equipment. See “Examining and Caring for the Flamepaths” in Chapter 5.
Replace parts immediately if damage or wear is apparent. Contact AMETEK if there is any
doubt about the integrity of any flamepath.
!
WARNING
!
WARNING
!
WARNING
!
WARNING