4-48 | Model 900 ADA / Model 930 Sulfur Pit Analyzers
Manual Start of Auto-Zero
A manual start of Auto-Zero resets the AZInt Timer to its initial value.
The start of the next Auto-Zero will be timed from the beginning of the
manual start.
Example:
The AZInt Timer is set at 2 hours, the time remaining until the next timed
Auto-Calibration is 30 minutes, and the Auto-Calibration is started manu-
ally. In this case, the AZInt Timer will be reset, and the time remaining
until the next timed Auto-Zero remains 2 hours.
Pressing Esc at any time during this procedure will abort the func-
tion and return to CAL mode normal display.
To manually start the Auto-Zero:
1. Press F1 0 (CAL mode) to display the timer duration.
2. Press Ent. The software prompts “Auto/Zero?”.
3. To enable the Auto-Zero, press Ent.
Timer0 will begin to count down to ‘0’. The analyzer Zero Offset is ad-
justed automatically based on the average of the readings during the
IntTime duration (or the last 15 seconds of the countdown if IntTime
= 0 seconds).
Or, press Esc to abort the Auto-Zero; the Zero Offset is not adjusted.
Upon completion of the sequence, enter another command, or press Esc
to return to RUN mode normal display.
Timed Start of Auto-Zero
The analyzer automatically initiates the Auto-Zero on a timed basis. The
AZInt timer sets the time interval between Zeroes.